The current measurement has 2 measurement positions
on the rotary switch: A and mA .
The A has a 8.0000A and 10.000A range, with
autoranging;the mA has a 80.000mA and 800.00mA
range, with auto ranging.
To measure current, connect the Meter as follows:
1. Turn off power to the circuit. Discharge all high-
voltage capacitors.
2. Insert the red test lead into the A or mA terminal
and the black test lead into the COM terminal. Use
the A terminal and A range if the current value to
be tested is an unknown.
3. Set the rotary switch to A or mA .
4. The Meter defaults to DC current measurement
mode. To toggle between DC and AC current
measurement function, press .
5. Break the current path to be tested. Connect the
red test lead to the more positive side of the break
and the black test lead to the more negative side
of the break.
6. Turn on power to the circuit.
The measured value shows on the display.
The AC current value displayed is a True RMS value.
G. Frequency & Duty Cycle Measurement
(see figure 11)
Frequency measurement
The measurement ranges are: 1kHz, 10kHz, 100kHz
and 1MHz. The maximum measurement range is 1MHz.
To measure frequency, connect the Meter as follows:
1. Insert the red test lead into the terminal
and the black test lead into the COM terminal.
2. Set the rotary switch to any measurement position
of V , V , mV , A , mA .
3. Press the Hz button once to display the frequency of
present signal while measure the AC voltage or
current.
4. Connect the test leads across with the object being
measured.
The measured value shows on the display.
Note
Press the Hz button in the MAX/MIN, HOLD mode,
it exits these modes.
12
Although it is possible to select frequency and duty
cycle
measurement mode at resistance, continuity and
diodes ranges, these functions cannot be used.
It is recommended to use mV range which has
the highest sensitivity.
To obtain a stable reading when measuring input
scope 30V rms frequency signal, set the rotary
switch to V or V .
When frequency measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test, and remove the testing
leads away from the input terminals of the Meter.
The attenuate amplitude of signal is different from
different measuring positions and ranges, therefore
the required input amplitude is different when
measuring frequency at different measuring positions
and ranges.
Duty Cycle Measurement
The duty cycle measurement range is 1% ~ 99%.
To measure duty cycle:
1. Set up the Meter to measure frequency.
2. To select duty cycle, press Hz again (or until the %
symbol is shown on the display).
3. Connect the test leads across the object being
measured.The measured value shows on the display.
Note
Press the Hz button in the MAX/MIN, HOLD mode,
it exits these modes.
The attenuate amplitude of signal is different from
different measuring positions and ranges, therefore
the required input amplitude is different when
measuring frequency at different measuring positions
and ranges.
Although it is possible to select frequency and duty
cycle measurement mode at resistance, continuity
and diodes ranges, these functions cannot be used.
It is recommended to use DCV 8V range.
The polarity of trigger slope is positive.
When duty cycle measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test, and remove the testing
leads away from the input terminals of the Meter.
MAX MIN Recording Mode(1) (see figure 12) MAX MIN Recording Mode(2)
MAX/MIN recording mode applied to all measurement
function except at conductance measurement function.
MAX/MIN recording mode captures and stores the
maximum and minimum input value detected, and
calculates a running average of all readings taken.
In the MAX/MIN mode, when the Meter detects an input
that is below the recorded minimum or above the recorded
maximum, the Meter beeps and records the new minimum
(MIN) or maximum (MAX) value.
In the MAX/MIN mode, you can also display a running
average of al readings taken since the MAX/MIN mode
was activated. Displaying the average reading allows
you to "smooth" a rapidly changing input by displaying
a stable reading on the digital display.
The MAX/MIN mode has a pre-set response time of
100ms. A shorter response time captures shorter events,
but with decreased accuracy. If you want 1 second
response time, that can be selected after power on the
Meter.
To use the MAX/MIN mode as follows:
Enter the
MAX/MIN
mode
Make sure that the Meter is in the desired
measurement function and range.
Press MAX/MIN to enter MAX/MIN recording
mode. The present reading is displayed, and
the Meter is locked in the present range, and
REC is displayed.
Each time a new low or high value is
recorded, the Meter beeps and displays a
new reading.
In MAX/MIN mode, Sleep Mode function
will be disabled
Press HOLD to stop recording, the recorded
values will not erase; Press HOLD again to
resume recording.
To select
response time
The pre-set response time is 100ms.
Press to select 1ms response time in
MAX/MIN mode. In 1ms response time, it
steps through on the sequence of MAX and
MIN only.
At the frequency measurement mode, it
should be 1second response time.
With the rotary switch in the OFF position,
press and hold down MAX/MIN while turning
on the rotary switch to select 1 second high
accuracy response time..
Press MAX/MIN to step through on the
sequence of present readings, high (MAX),
low (MIN) and average (AVG) readings.
To stop and
resume
recording
without erasing
stored value
To view the MAX,
MIN, AVG or
present reading
Press MAX/MIN for over 1 second to exit the
mode. Recorded values are erased and the
Meter stays in the selected range.
To exit the
MAX/MIN mode
Peak Test
PEAK MAX MIN mode is an additional function of .
Pressing to toggle between 100ms response time
and 1ms response time. This function can only be enabled
when the Meter is under MAX/MIN, HOLD mode.
To enter PEAK MAX MIN mode:
Press to enter PEAK MAX MIN mode (1ms
response time), it steps through the sequence of
MAX and MIN reading only.
Press HOLD under 1ms recording mode to stop
recording without erasing stored values. Press
HOLD again to resume recording.
12
Operation of Hold Mode
The HOLD mode is applicable to all measurement
functions. With the Auto Hold function, it automatically
holds the present reading on the display, freeing you to
concentrate on the placement of the probes when working
in dangerous or difficult situations.
To avoid possibility of electric shock, do not use
HOLD mode to determine if circuits are without power.
The HOLD mode will not capture unstable or noisy
readings.
Press HOLD to enter or exit the HOLD mode:
Press HOLD to enter HOLD mode, the Meter beeps.
Press HOLD again or RANGE or turn the rotary
switch to exit HOLD mode, the Meter beeps.
In HOLD mode, DATA HOLD is displayed on the
LCD.
When measure resistance higher than 12.5MΩ ,
conductance value can be displayed by pressing
HOLD. To switch to resistance value, press HOLD
again.
Conditions to enter Auto Hold Mode:
When input signal is larger than AC/DC 100mV, Auto
Hold mode will be enabled automatically.
When the change of signal three times continuously
within 3 digits, it will hold the last value; the Meter
beeps.
Considers every value as the base value. When the
Meter detects an input which is above the base value
300 digits, the Meter beeps, and holds and displays
the value as the new base value.
Operation of Auto Hold Mode
Auto Hold means you no need to press HOLD, the readings
on the LCD will be hold automatically.